Operating the Temperature Controllers (Large Cell)
There are identical control setups for each cell. These consist of a programmable temperature controller and a limit controller. The limit controller shuts down the heating supply if the equipment temperature exceeds the limit set-point chosen or if there is an electrical short in a thermocouple or thermocouple wire. The procedure for turning on the system, setting up a temperature program, and starting up the system is as follows:
| Table III. Example for Setting Controller |
| Display |
Setting |
Description |
| Pn |
1 |
Profile number |
| RS |
1 |
Run segment |
| Rt |
3.00 |
Ramp time in hours |
| SP |
250 |
Set point in Celsius |
| ST |
24.00 |
Soak time in hours |
| PLCt |
1 |
Profile loop count |
| Dhru |
1 |
|
| Dhrd |
0.0 |
|
| Pend |
0 |
Program end |
| Table IV. Profile Number Reservations |
| Profile # |
Large Rocker |
Small Rocker |
| 1 |
|
|
| 2 |
Yoonkyung |
Yoonkyung |
| 3 |
|
|
| 4 |
|
|
| 5 |
|
|
| 6 |
|
Raj |
| 7 |
|
|
| 8 |
Hardas-OpTest |
Hardas-OpTest |
| |
|
|
- Return to the top of the software tree by pressing the "up" key while one of the parameter names (Pn, rS, rt, etc.) is being displayed.
- Scroll through the parameters again to make sure that they are set correctly.
- Adjust the limit set-point on the limit controller while holding the toggle switch in the direction of the "Set II" setting.
- Reset the contacts for the heating circuit by pressing the reed "start" button to the above right of the controllers.
- Scroll to Pn on the temperature controller (where n=your profile #) and press the run/hold key.
- Be sure to stay around long enough to see that the equipment is indeed heating up. You should notice a change in temperature of at least 5 °C on each controller before assuming that everything is set properly.
- After the total run time (ramp time plus soak time), the heat will turn off. Be sure to let the temperatures cool down sufficiently before attempting to open or depressurize a reaction vessel. Also remember to turn off the power to the equipment and instruments in the circuit breaker box when you are finished.
- If you need additional information on operating the temperature controllers, please consult the temperature controller manuals in another section of this lab manual.